POP-7™ (384) Polymer for 3500 Dx/3500xL Dx Genetic Analyzers 3500 Dx Series POP-7™ and POP-6™ are separation matrices for the 3500 Dx/3500xL Dx Genetic Analyzers. They are conveniently offered in easy-to-use pouch packages with either 960 samples (maximum) or 384 samples (maximum) size formats. Each polymer is optimized for specific applications. Easy to Use Polymer is pre-formulated in a single use pouch, which not only saves time but also ensures reliability and consistency. Radio Frequency Identification (RFID) label The RFID encodes easy-to-track important information such as part/lot number, samples remaining and expiry dates (displayed on software dashboard) Robust Formulation Polymers are formulated for a wide range of applications. Reproducible Results POP™ polymers dynamically coat the capillary wall to control electro-osmotic flow. Their definite quality and uniform consistency eliminate guesswork and help to ensure reproducibility. POP Polymers are specifically formulated to separate DNA fragments of a known size range at a Accuracy: 98.5% up to 900 bases (50 cm capillary), 98.5% up to 700 bases (36 cm capillary) Platform: 3500xL Dx Genetic Analyzer, 3500 Dx Genetic Analyzer Run Time: 2 hrs for 50 cm capillary (long read), 1 hr for 36 cm capillary (std. seq.) Product Size: 384 samples Capillary Length: 50 cm, 36 cm Recommended Applications: Standard Sequencing, Long-Read Sequencing, Short Read Sequencing, Fragment Analysis (Denaturing) On-Instrument Life⁄Usage: Contains enough polymer for 384 runs; polymer is stable on instrument for 7 days Shipping Condition: Wet Ice Regulatory Statement: For In Vitro Diagnostic Use. |